Tektronix CSA8000B Manuale Utente Pagina 48

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 354
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 47
Incoming Inspection
1-30
CSA8000B & TDS8000B User Manual
NOTE. At some temperatures, there may be extraneous data points after the first
half cycle when viewing the front-panel Internal Clock output (as is done in this
step). This behavior may also occur when viewing multiple cycles in TDR mode.
In both cases, this behavior is normal.
H Rotating the Horizontal SCALE knob clockwise expands the waveform
on-screen (more horizontal divisions per waveform period), counter-
clockwise rotation contracts it, and returning the horizontal scale to
1 s/div returns the period to about five divisions. Leave the time base
set to 1 s/div.
H The horizontal POSITION knob positions the signal left and right
on-screen when rotated.
NOTE. The signal will not move past the minimum position setting.
Control bar
Vertical scale
setting
Horizontal
scale setting
Internal clock
signal
s
Figure 1- 13: Main time base verification