Performance Tests
1–74
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Digitizing Oscilloscope
50 cable
Signal Generator
Output
Figure 1–31: 60 Hz Rejection test hookup
H Adjust the signal generator for three vertical divisions of 60 Hz
signal. See Figure 1–32. The signal will not be triggered. That is, it
will run free.
Figure 1–32: 60 Hz Rejection test setup signal
c. Check 60 Hz rejection:
H Use the keypad to set the horizontal scale (/div) to 50 s (press 50,
SHIFT, m, then ENTER).
Vedere la pagina 89