Tektronix Thermometer P7225 Manuale Utente Pagina 28

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 56
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 27
Functional Check
16
P7225 2.5 GHz 10X Active Probe User Manual
3. Using a ground lead and SMT KlipChip, connect the probe
ground to the instrument ground on the host instrument.
4. Using a standard tip, hold the probe to the center conductor of the
PROBE COMPENSATION output on the host instrument.
5. Press AUTOSET, or adj ust the host instrument to display a stable
calibration waveform. (You may need to adjust the probe offset to
display the waveform.)
NOTE. Now is a good time to perform a probe calibration routine
from your instrument. Refer to the online help of your host instrument
for calibration routine procedures.
6. Disconnect the probe tip from t he host instrument, and ground the
probe tip. (Connect the KlipChip to the probe tip.)
7. Set the probe offset to 0.0 V to m ake the host instrument display
at the ground reference.
8. Adjust the V/div to keep the waveform on the screen.
Vedere la pagina 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 55 56

Commenti su questo manuale

Nessun commento