Tektronix AWG610 Manuale Utente Pagina 385

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 466
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 384
Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-25

DMM
50 W BNCĂcoaxial cable
InputĂconnector
+ĂBNCĆtoĆdual banana adapter
+Ă50 W BNCĂcoaxialĂcable
AWG610 Arbitrary Waveform Generator
CH1ĂoutputĂconnector
+ĂSMA(Ma)ĆBNC(Fe) adapter
+ĂBNC 50 W Ăprecision terminator
+Ă50 WĂBNC coaxial cable
Figure B-9: Amplitude accuracy initial test hookup
2. Set the DMM controls as follows:
Mode . ......................... VDC
Range ......................... Auto
Input .......................... Front
3. Follow the substeps below to set the AWG610 Arbitrary Waveform
Generator controls and to select the sequence file:
a. Push UTILITY (front-panel)System (bottom)Factory Reset
(side)OK (side).
b. Push SETUP (front-panel)Run Mode (bottom)Enhanced (side).
The AWG610 Arbitrary Waveform Generator is set to enhanced mode.
c. Load the AMP1.SEQ file.
Refer to Loading Files on page B–10 for file loading procedures.
4. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connector are on.