Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual

Sweep ..................... 1ms/div
Trigger
Source . .................... CH1
Coupling .................... DC
Slope ...................... Positive
Level ...................... 0.2V
Mode . ..................... Auto
3. Push UTILITY (front-panel)! System (bottom)!Factory Reset
(side)!OK (side).
4. Load the MODE.WFM file.
Refer to Loading Files on page B–10 for file loading procedures.
5. Push SETUP (front-panel)!Run Mode (bottom)!Triggered (side).
6. Push Horizontal (bottom)!Clock (side)!1, 0, 0, M (SHIFT+7).
The clock is set to 100.0 MS/s.
7. Follow the substeps to set the trigger interval:
a. Push SETUP (front-panel)!Trigger (bottom)!Source (side)!
Internal (side).
b. Push the Interval side button.
c. Push 1 and m (SHIFT+9) keys in this order or turn the general pur-
pose knob to set the trigger interval to 1 ms.
The numeric value of 1m is entered.
8. Push the RUN and CH1 OUT buttons.
The LEDs above the RUN button and CH1 output connectors are on.
9. Verify that there is a single sinewave cycle displayed at 1 ms intervals.
10. Push 2 and m (SHIFT+9) keys in this order or turn the general pur-
pose knob to set the trigger interval to 2 ms.
The trigger interval is changed to a value of 2 ms.
11. Verify that the period between the waveform displayed on the oscilloscope is
two horizontal divisions.
12. Push the CH1 OUT button to turn off the CH1 LED.