Tektronix 784A Manuale Utente Pagina 120

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 119
Performance Tests
4–44
TDS 684A, TDS 744A, & TDS 784A Service Manual
Check long–term sample
rates and delay time
accuracies against limits.
4
Align the trigger T to the
center graticule line.
1
Set
horizontal
mode.
2
Set horizontal scale
and delayed time.
3
Figure 4–12: Measurement of Accuracy — Long-Term and Delay Time
3. Disconnect the hookup: Disconnect the cable from the generator output at
the input connector of CH 1.
Trigger System Checks
These procedures check those characteristics that relate to the Main and Delayed
trigger systems and are listed as checked in Specifications.
Equipment
Required
One medium-frequency leveled sine wave generator (Item 11)
One 10X attenuator (Item 1)
One 50 , precision coaxial cable (Item 5)
Prerequisites See page 4–15
Check Accuracy (Time) for
Pulse-Glitch or
Pulse-Width Triggering
Vedere la pagina 119
1 2 ... 115 116 117 118 119 120 121 122 123 124 125 ... 293 294

Commenti su questo manuale

Nessun commento