Tektronix 784A Manuale Utente Pagina 64

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 63
Operating Instructions
2–10
TDS 684A, TDS 744A, & TDS 784A Service Manual
2. Select the waveform source(s). Position, scale, and set trigger level for
waveform source(s) directly from the front panel. The steps that follow
provide examples of possible control selections.
123
65
7
8
4
a. Input waveforms into these channels (7). Example: CH 1.
b. Push any channel’s button (8) to display its waveform. The last channel
turned on determines which waveform is positioned and scaled. The
indicator above the channel last selected is lighted.
Example: Push CH 1; then CH 2.
c. Vertically (1) and horizontally (2) scale and position the waveform(s)
selected. Example: Set the scale to 100 mV/div and center the waveform
on screen.
d. Stop and start acquiring waveforms (3). Example: Push RUN/STOP if
not acquiring.
Vedere la pagina 63
1 2 ... 59 60 61 62 63 64 65 66 67 68 69 ... 293 294

Commenti su questo manuale

Nessun commento