Setting SONETSDH Receive Parameters
CTS 710 SONET Test Set User Manual
3–119
To drop data from the User Channel:
1. Connect an external protocol analyzer to the CTS 710 using the
Overhead Add/Drop Port on the rear panel of the CTS 710.
2. Configure the CTS 710 to drop the User Channel data byte:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
RECEIVE PATH
OVERHEAD
External Drop
(see Figure 3–43)
None
F2
H Select None if you do not want to drop the User Channel data.
H Select F2 to drop the User Channel byte.
Figure 3–43: Dropping the User Channel Data Byte
Setting SONETSDH Receive Parameters
CTS 710 SONET Test Set User Manual
3–119
To drop data from the User Channel:
1. Connect an external protocol analyzer to the CTS 710 using the
Overhead Add/Drop Port on the rear panel of the CTS 710.
2. Configure the CTS 710 to drop the User Channel data byte:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
RECEIVE PATH
OVERHEAD
External Drop
(see Figure 3–43)
None
F2
H Select None if you do not want to drop the User Channel data.
H Select F2 to drop the User Channel byte.
Figure 3–43: Dropping the User Channel Data Byte