Tektronix DPO4000 Manuale Utente Pagina 279

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 585
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 278
Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4PWR application module.
Group
Power
Syntax
POWer:INDICators {OFF|ON|0|1}
POWer:INDICators?
Arguments
OFF or 0 turns off the measurement indicators.
ON or 1 turns on the measurement indicators.
Examples
POWER:INDICATORS 1 turns on the indicator.
POWER:INDICATORS? might return :POWER:INDICators 0 indicating that
the indicator is off.
POWer:MODulation:SOUrce
Sets or returns the source waveform for modulation tests. The voltage source
waveform is specied using the POWer:VOLTAGESOurce command and the
current waveform is specied using the POWer:CURRENTSOurce command.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:MODulation:SOUrce {VOLTage|CURRe nt}
POWer:MODulation:SOUrce?
Arguments
VOLTage species voltage source waveform for modulation tests.
CURRent species current source waveform for modulation tests.
Examples
POWER:MODULATION:SOURCE CURRent sets the modulation source to Current.
POWER:MODULATION:SOURCE? might return :POWER:MODULATION:SO URCE
VOLT
indicating that the modulation source is set to Voltage.
MSO4000 and DPO4000 Series Programmer Manual 2-269