Reference
TDS1000/2000-Series Digital Oscilloscope User Manual
111
Key Points
Self Calibration. The self calibration routine optimizes the oscilloscope
accuracy for the ambient temperature. For maximum accuracy, you
should perform a self cal i f the ambient temperature changes by 5° C
or more. Follow the di rections on the screen.
System Status
Selecting System Status from the Utility Menu displays the menus
available for obtaining a list of control settings for each group of
oscilloscope controls.
Push any front-panel m enu button to remove the status screen.
Options Comments
Horizontal Lists horizontal parameters of channels
Vertical Lists vertical parameters of channels
Trigger Lists trigger parameters
Misc Lists the model of the oscilloscope and the software version
number
If the TDS2CMA module is installed, lists values of the
communications parameters