TDS1000/2000-Series Digital Oscilloscope User Manual
75
Sample. Use Sample acquisition mode to acquire 2500 points and
display them at the SEC/DIV setting. Sample mode is the default
mode.
12345678910
Sample acquisition intervals (2500)
Sample mode acquires a single sample point in each interval.
Sample points
The maximum sample rate is 1 GS/s for oscilloscope models with a
bandwidth of 60 MHz or 100 MHz and 2 GS/s for the 200 MHz
models. At 100 ns and faster settings, this sample rate does not
acquire 2500 points. In thi s case, a Digital Signal Processor
interpolates point s between the sampled points to make a full 2500
point waveform record.