Getting Started
TDS1000/2000-Series Digital Oscilloscope User Manual
5
Functional Check
Perform this quick functiona l c heck to verify that your osc illoscope
is operating correctly.
PASSED
ON
/
OFF
button
1. Power on the oscilloscope.
Wait until the display shows that all
power-on tests passed. Push the DEFAULT
SETUP button. The default Probe option
attenuation setting is 10X.
CH 1
PROBE COMP
2. Set the switch to 10X on the P2200 probe
and connect the probe to channel 1 on the
oscilloscope. To do this, align the slot in
the probe connector with the key on the
CH 1 BNC, push to connect, and t wist to
the right to lock the probe in place.
Connect the probe tip and reference lead
to the PROBE COMP connectors.
3. Push the AUTOSET button. Within a few
seconds, you should see a square wave i n
the display of about 5 V peak-to-peak at
1kHz.
Push the CH 1 MENU button twice to
remove channel 1, push the CH 2 MENU
button to display channel 2, repeat steps 2
and 3. For 4-channel models, repe at for
CH 3 and CH 4.